Skip to article frontmatterSkip to article content
Site not loading correctly?

This may be due to an incorrect BASE_URL configuration. See the MyST Documentation for reference.

Function generator: Tektronix AFG 3021B/C

Introduction

The lab is equipped with the Tektronix AFG 3021B/C arbitrary function generator. This appendix will briefly discuss the operation of this device. For more information, please consult the user manual that is available in the lab room.

Overview

A diagram of the AFG 3021B/C is depicted in Figure 1. The buttons on the devices are grouped, based on functionality.

The Tektronix AFG 3021B/C Arbitrary Function Generator.

Figure 1:The Tektronix AFG 3021B/C Arbitrary Function Generator.

Run Mode

On the top, there are four buttons grouped as Run Mode. These perform the following functions:

Function

Directly left from the Run Mode group is the Function selection. These allow quick a choice between the following waveforms:

Signal Setup

Under the Run Mode buttons are buttons to set up the selected waveform. Each button has two (related) functions, repeatedly pressing the button will alternate between the functions.

When the Pulse waveform is selected, two additional buttons are available:

Pressing one of these buttons highlights the corresponding setting on the screen. The value can now be changed by either typing in the numerical value on the keypad and pressing Enter to confirm, or by using the general purpose rotary knob. The arrow keys directly under the general purpose knob move the cursor and thus the digit changes when turning the general purpose knob.

The screen shows also shows a preview of the signal annotated with all the important voltage levels and frequencies. These levels are only valid when the output impedance is properly set up, as described in Section \ref{sec:app.function_gen.output_imp}.

Additional Signal Setup

The buttons directly next to the screen are context-sensitive and are used to modify additional signal parameters. Changing the parameters is identical as described in Section \ref{sec:app.function_gen.signal_setup}.

Output and Output Impedance

The output of the generator is the leftmost coax connector labeled Output. The amplitude settings on the generator are only valid when the value of the load impedance is properly configured. By default, the load impedance is assumed to be 50 Ω. The input impedance of, for example the oscilloscope, op-amp inputs and almost any digital IC input is very high. In order to have the actual voltage levels match the settings on the function generator, the output impedance has to be changed to high impedance. You can use the following steps to set the output impedance to high impedance.

The lower button can be used to return to the main screen.

Note that the output settings are only valid for a fixed load impedance; if the load is highly variable, make sure to verify the actual output on an oscilloscope or use a buffer circuit.

The output signal is enabled or disabled by pushing the On button in the block Channel just above the output connector.